Abstract:
By accurately measuring the position of the peaks in the ion mass spectrum of residual gases, it is found that the peaks are associated with hydrocarbon ions. The “mass defect” spectrum is close to that obtained analytically. The deviations observed are related to the presence of the $^{13}$C isotope and nonsaturation of bonds in the ions recorded. The experimental data obtained may be used as a reference in impurity microanalysis, in calibrating the mass scale of mass spectrometers, and in checking the theory of mass spectrum analysis.