RUS  ENG
Full version
JOURNALS // Zhurnal Tekhnicheskoi Fiziki // Archive

Zhurnal Tekhnicheskoi Fiziki, 2020 Volume 90, Issue 9, Pages 1475–1480 (Mi jtf5202)

III International Conference Physics -- Life Sciences
Radiospectrometric methods in sciences of life

Optimization of parameters of RF pulse sequences for MRI investigations in the presence of magnetic nanoparticles

Yu. V. Bogacheva, A. V. Nikitinaa, V. V. Frolovb, Ya. Yu. Marchenkoa, B. P. Nikolaeva

a Saint Petersburg Electrotechnical University "LETI"
b Saint Petersburg State University

Abstract: The influence of parameters of various pulsed radio frequency (RF) sequences on the contrast of magnetic resonance (MR) images is investigated. A program for modeling of magnetic resonance imaging (MRI) tomograms and determining the optimal values of pulse RF sequences parameters to achieve the best contrast of MR images has been developed. This program has been tested for MRI studies using negative contrast agents based on magnetic iron oxide nanoparticles and positive contrast agents based on gadolinium. Their influence on MR signal intensity depending on concentration is analyzed. Recommendations on the choice of optimal values of RF sequences parameters for MRI diagnostics using contrasting agents are developed.

Keywords: MRI, radio frequency pulse sequence, magnetic nanoparticles, theranostic.

Received: 19.12.2019
Revised: 19.12.2019
Accepted: 17.02.2020

DOI: 10.21883/JTF.2020.09.49678.419-19


 English version:
Technical Physics, 2020, 65:9, 1416–1420

Bibliographic databases:


© Steklov Math. Inst. of RAS, 2024