Zhurnal Tekhnicheskoi Fiziki, 2018 Volume 88, Issue 1, Pages 141–146 (Mi jtf6034)

This article is cited in 1 paper

Experimental instruments and technique

New possibilities of positron-emission tomography

A. N. Volobuev

Samara State Medical University, Department of medical and biological physics, Samara

Abstract: The reasons for the emergence of the angular distribution of photons generated as a result of annihilation of an electron and a positron in a positron-emission tomograph are investigated. It is shown that the angular distribution of the radiation intensity (i.e., the probability of photon emission at different angles) is a consequence of the Doppler effect in the center-of-mass reference system of the electron and the positron. In the reference frame attached to the electron, the angular distribution of the number of emitted photons does not exists but is replaced by the Doppler shift of the frequency of photons. The results obtained in this study make it possible to extend the potentialities of the positron-emission tomograph in the diagnostics of diseases and to obtain additional mechanical characteristics of human tissues, such as density and viscosity.

Received: 25.01.2017

DOI: 10.21883/JTF.2018.01.45498.2178


 English version:
Technical Physics, 2018, 63:1, 141–147