Zhurnal Tekhnicheskoi Fiziki, 2017 Volume 87, Issue 1, Pages 130–135 (Mi jtf6350)

This article is cited in 2 papers

Experimental instruments and technique

Some principles in choosing parameters of magnetic resonance tomographs

A. N. Volobuev

Samara State Medical University

Abstract: The problem of amplifying the signal that ensures the visualization of internal organs in the magnetic resonance tomograph due to the optimal selection of some of its parameters has been considered. The operating principle of the tomograph has been analyzed. The relation between the angle of the magnetic moment precession in hydrogen nuclei in an organism, the frequency of the ac magnetic field exciting this precession, and the constant magnetic field used has been determined using quantum-mechanical concepts. This relation makes it possible to determine the optimal parameters for tomograph operation.

Received: 09.02.2016

DOI: 10.21883/JTF.2017.01.44029.1755


 English version:
Technical Physics, 2017, 62:1, 157–163