Mendeleev Commun., 2016 Volume 26, Issue 1, Pages 24–25 (Mi mendc2097)

This article is cited in 4 papers

Communications

Sensitivity of magnetic resonance imaging based on the detection of 19F NMR signals

M. V. Gulyaeva, N. V. Anisimova, N. E. Ustyuzhaninab, A. S. Dmitrenokb, A. S. Shashkovb, D. V. Yashunskyb, L. L. Gervitsc, Yu. A. Pirogova, N. E. Nifantievb

a Department of Fundamental Medicine, M.V. Lomonosov Moscow State University, Moscow, Russian Federation
b N.D. Zelinsky Institute of Organic Chemistry, Russian Academy of Sciences, Moscow, Russian Federation
c A.N. Nesmeyanov Institute of Organoelement Compounds of Russian Academy of Sciences, Moscow, Russian Federation

Abstract: The evaluation of the sensitivity of magnetic resonance tomography with a magnetic field of 7 T based on the detection of 19F NMR signals was performed with the use of the solutions of trifluoroacetic acid, 1,1,1,3,3,3-hexafluoroisopropanol and sym-nonafluoroneopentanol, whose 19F NMR spectra consist of single line signals. The scanning of a 12×12cm zone with a resolution of 1.2mm and a cut thickness of 4mm for 10min gave reliable images if the local concentration of 19F atoms in the test zone exceeded 0.3–0.6%.

Language: English

DOI: 10.1016/j.mencom.2016.01.010