Research of spectral characteristics of optical absorption gas analyzer for freon monitoring in air medium

L. A. Konopelkoa, E. P. Tyurikovaa, Y. N. Snytkob

a St. Petersburg National Research University of Information Technologies, Mechanics and Optics
b SPA "Analitpribor", Smolensk

Abstract: The optical scheme of optical absorption gas analyzer which optimized for measuring the freons content, is presented. The radiation source with a given maximum of spectral density of radiant emittance in the range from 8 to 10 $\mu$m, with a reference channel of 9.5 $\mu$m and a measurement threshold of 1 ppm is developed and studied. The design of the interference filter system is developed.

Keywords: gas analyzer, freon, absorption spectroscopy, pyroelectric detector.

Received: 19.11.2019
Revised: 05.02.2020
Accepted: 07.02.2020

DOI: 10.21883/OS.2020.05.49329.308-19


 English version:
Optics and Spectroscopy, 2020, 128:5, 678–685