Intracavity spectroscopy with Ti3+:Al2O3 lasers pumped by short pulses

V. B. Andrienko, Yu. N. Parkhomenko, O. N. Galkin

Sonar Scientific-Engineering Centre, Academy of Sciences of Ukraine, Kiev

Abstract: An effective and simple intracavity laser spectroscopy method is proposed and investigated. This method is based on the feasibility of a considerable increase in the duration of lasing and of control of this duration within wide limits in systems with a longitudinal relaxation time considerably longer than the duration of pumping. The results of an experimental investigation show that the use of a Ti3+:Al2O3 laser can enhance additionally (by a factor of 102, which may be increased to ~103) the sensitivity of this method, compared with extracavity absorption spectroscopy. The sensitivity control interval was 10–4 — 10–5 cm–1. A distinguishing feature of the method is the use of a special algorithm for the processing of signals in order to eliminate the influence of strong fluctuations of the laser radiation energy (duration).

PACS: 42.62.Fi, 07.60.Rd

Received: 01.01.1996


 English version:
Quantum Electronics, 1996, 26:5, 462–463