RUS  ENG
Full version
JOURNALS // Kvantovaya Elektronika // Archive

Kvantovaya Elektronika, 1977 Volume 4, Number 1, Pages 148–150 (Mi qe8819)

This article is cited in 1 paper

Brief Communications

Intracavity neodymium-glass laser spectrometer with a dispersive resonator

L. N. Sinitsa


Abstract: A description is given of a neodymium-glass laser spectrometer with a dispersive resonator. This spectrometer operates in the 9230–9520 cm–1 range with a resolution between 0.1 and 0.02 cm–1 and can be used to detect absorption characterized by a coefficient of 10–7 cm–1. This spectrometer was used to obtain a low-resolution record of the P branch of the 1203 band, the 0403 band, and the "hot" 1313→0110 transition in carbon dioxide gas as well as the absorption spectra of atmospheric air and water vapor. A high-resolution record was obtained of the edge of the 0403 band and the distances between the lines were measured.

UDC: 535.338.332:621.378.325

PACS: 07.65.-b, 42.60.Kg

Received: 24.11.1975


 English version:
Soviet Journal of Quantum Electronics, 1977, 7:1, 77–79